home product info customer service search 4 port application 6 port application 8 port application 10 port application

home

links

Back to Cheminert injectors and valves
product heading



VICI heading




C5 valve

HPLC
Stream selector

1/16" Valco ZDV fittings, 0.40 mm ports (.016")
Model C5

The Cheminert® Model C5 HPLC stream selector valve selects and isolates one of 4 to 10 streams, with the remainder dead-ended.

The C5 valve is the backbone of the Cheminert HPLC column selector system, which includes two stream selection valves mounted on a single microelectric actuator.

An optional flowpath, Model C5F , the flow-through version, is similar to the C5 but its non-selected streams continue flowing through individual outlets. 3, 4, and 5 positions are available.


flowpath   Dead-ended stream selector
 
See also:

Optional flow-through flowpath

Specifications / Product numbers

A complete valve product number includes information about the specific valve, actuator, and mounting hardware. Locate the valve with desired specifications in the chart below and select it to continue to the next step in arriving at the correct product number.
Jump to the valve selection chart.


Actuation

Valves can be actuated manually or automated with pneumatic or electric actuators. Read more about actuators.

HPLC valves rarely require a standoff. Contact the factory if you need one.


Options

Consult the factory for product numbers and pricing for the following options:
• Optional flow-through flowpath (read more)
• Hastelloy C stators
• 2", 3", 4", and 6" standoffs
• Optional 0.25 mm (.010") and 150 µm (.006") bores available
• Optional 0.75 mm (.030") bore for Prep HPLC available

Stream selector selection chart

For specific product numbers, select a valve specification below.

 Specs   Stator   Rotor   
 50°C max
 5000 psi liq 
 PAEK   Valcon E   Select
 75°C max *
 5000 psi liq 
 N60 stainless   Valcon H   Select
 Titanium   Valcon H   Select
* Electric actuators are rated at 40°C; if the valve will be heated above 40°C,
* a standoff is required between the valve and actuator.
to top

Further reference

Which valve do I need?
Properties of:
   Metals
   Polymers
   Rotor materials

to top

Optional flowpath

Model C5F, the flow-through version, is similar to the C5 but its non-selected streams continue flowing through individual outlets. 3, 4, and 5 positions are available.
 
Consult the factory for C5F prices and information.

  flowpath
to top

Home | Products | Customer service | Contact | About VICI | Site map | Site index

VICI®, Valco®, and Cheminert® are registered trademarks of Valco Instruments Co. Inc. and VICI AG
© 2006 VICI